I need two inputs for my project. One thermocouple board works, the other doesn't. I swapped out the wiring and the problem stayed with the BOARD. I changed the sensors and the problem stayed with the BOARD. One of the two BOARDS is defective. Also, one of the thermocouple wires isn't properly connected to the spade connector, when I move it the signal goes to either "NAN" (no signal) or the maximum, indicating a complete burnout of the scale. The spade tips have been removed and connected directly to the terminals on the board and the thermocouple has a stable signal. I would like to change my rating if the supplier sends a replacement board and sensor for replacement.